Abstract
The Use of a Differential Mobility Analyzer to Perform Aerosol Size Spectrum Measurements Requires an Inversion Method to Go from the Measured Sensor Responses to the Desired Size Spectrum Information. Here We Present a Linear Inversion Method that Can Be Run on a Microcomputer or a Small Minicomputer. It Does Not Use Some of the Approximations Made in the Techniques Currently Available and Hence Gives Better Inversion Accuracy. the Method Shows Good Immunity to Both Random and Systematic Experimental Error. It is Applied to Numerous Test Case Aerosols. © 1983 Elsevier Science Publishing Co., Inc.
Recommended Citation
D. E. Hagen and D. J. Alofs, "Linear Inversion Method to Obtain Aerosol Size Distributions from Measurements with a Differential Mobility Analyzer," Aerosol Science and Technology, vol. 2, no. 4, pp. 465 - 475, Taylor and Francis Group; Taylor and Francis, Jan 1983.
The definitive version is available at https://doi.org/10.1080/02786828308958650
Department(s)
Mechanical and Aerospace Engineering
Second Department
Physics
International Standard Serial Number (ISSN)
1521-7388; 0278-6826
Document Type
Article - Journal
Document Version
Citation
File Type
text
Language(s)
English
Rights
© 2023 Taylor and Francis Group; Taylor and Francis, All rights reserved.
Publication Date
01 Jan 1983
Comments
National Science Foundation, Grant ATM79-24326