Masters Theses
Abstract
"The purpose of this research was to calibrate an instrument with which an accelerometer could be calibrated accurately. The instruments used were a piezoelectric shaker and a modified Michelson interferometer together with necessary electronic equipment. This equipment was used to calibrate a Kistler Model 808A accelerometer.
Weight was added to the shaker in the second part of this experiment in order to determine its effect on the amplitude and frequency of oscillation of the shaker.
Results show that this instrument may be used to calibrate an accelerometer quite accurately. The weight added to the shaker will effect only the amplitude of vibration and not the frequency of oscillation"--Abstract, page ii.
Advisor(s)
Edwards, Charles L., 1934-
Committee Member(s)
Rhea, L. G.
Sauer, Harry J., Jr., 1935-2008
Davidson, Robert F., 1911-1971
Department(s)
Mechanical and Aerospace Engineering
Degree Name
M.S. in Mechanical Engineering
Publisher
University of Missouri at Rolla
Publication Date
1966
Pagination
vii, 47 pages
Note about bibliography
Includes bibliographical references (pages 42-43).
Rights
© 1966 Nader Khorzad, All rights reserved.
Document Type
Thesis - Open Access
File Type
text
Language
English
Subject Headings
Accelerometers -- Design and constructionAccelerometers -- Calibration
Thesis Number
T 1941
Print OCLC #
5977622
Electronic OCLC #
906042365
Recommended Citation
Khorzad, Nader, "Calibration of an instrument to calibrate accelerometers" (1966). Masters Theses. 2962.