Abstract
Absolute detection efficiencies for singly and multiply charged positive ions have been measured for a channelplate and for two different channel electron multipliers (CEM). The efficiencies were measured for impact energies between approximately 0.25 and 25 keV and for ion masses ranging from 14 to 132. The maximum efficiencies were found to be the same for all ions investigated and were approximately 58% for a channelplate and 89% for the CEMs. For a channelplate it is shown that the detection efficiencies for heavier ions scale to a single curve if plotted versus the impact energy divided by the square root of the ion mass. Data taken from the literature imply that lighter ions scale differently. Polynomial fitting parameters to the present efficiency curves are provided.
Recommended Citation
M. Krems et al., "Channel Electron Multiplier and Channelplate Efficiencies for Detecting Positive Ions," Review of Scientific Instruments, American Institute of Physics (AIP), Sep 2005.
The definitive version is available at https://doi.org/10.1063/1.2052052
Department(s)
Physics
Sponsor(s)
National Science Foundation (U.S.)
United States. Department of Energy
Keywords and Phrases
Electron Multiplier Detectors; Positive Ions; Radiation Detection; Cations; Curve fitting
International Standard Serial Number (ISSN)
0034-6748
Document Type
Article - Journal
Document Version
Final Version
File Type
text
Language(s)
English
Rights
© 2005 American Institute of Physics (AIP), All rights reserved.
Publication Date
01 Sep 2005