Abstract
Properties of TE011 cylindrical, microwave cavities containing cylindrical samples of various radii and dielectric constants are calculated. The properties considered are the resonant frequency, quality factor (Q), relevant magnetic filling factor for spin transitions (ε), and a signal sensitivity factor (Qε) for a lossless sample. Sample sizes range from zero radius to full cavity radius with some experimental data on less than full length samples. The choice of dielectric constants ranges from one to sixteen. The data are presented in dimensionless form since they will be of use to other ESR experimentalists. It is shown that use of large samples is undesirable even if they are lossless. Furthermore, elongated cavities (D/L ratios less than one) are to be preferred over shortened cavities. © 1973 The American Institute of Physics.
Recommended Citation
K. A. Rages et al., "Properties Of Microwave Cavities Containing Magnetic Resonant Samples," Review of Scientific Instruments, vol. 44, no. 7, pp. 830 - 834, American Institute of Physics, Dec 1973.
The definitive version is available at https://doi.org/10.1063/1.1686258
Department(s)
Physics
International Standard Serial Number (ISSN)
0034-6748
Document Type
Article - Journal
Document Version
Final Version
File Type
text
Language(s)
English
Rights
© 2023 American Institute of Physics, All rights reserved.
Publication Date
01 Dec 1973